A Practical Guide to Instrumental Analysis
eBook - ePub

A Practical Guide to Instrumental Analysis

Erno Pungor, G. Horvai

  1. 288 páginas
  2. English
  3. ePUB (apto para móviles)
  4. Disponible en iOS y Android
eBook - ePub

A Practical Guide to Instrumental Analysis

Erno Pungor, G. Horvai

Detalles del libro
Vista previa del libro
Índice
Citas

Información del libro

A Practical Guide to Instrumental Analysis covers basic methods of instrumental analysis, including electroanalytical techniques, optical techniques, atomic spectroscopy, X-ray diffraction, thermoanalytical techniques, separation techniques, and flow analytical techniques. Each chapter provides a brief theoretical introduction followed by basic and special application experiments. This book is ideal for readers who need a knowledge of special techniques in order to use instrumental methods to conduct their own analytical tasks.

Preguntas frecuentes

¿Cómo cancelo mi suscripción?
Simplemente, dirígete a la sección ajustes de la cuenta y haz clic en «Cancelar suscripción». Así de sencillo. Después de cancelar tu suscripción, esta permanecerá activa el tiempo restante que hayas pagado. Obtén más información aquí.
¿Cómo descargo los libros?
Por el momento, todos nuestros libros ePub adaptables a dispositivos móviles se pueden descargar a través de la aplicación. La mayor parte de nuestros PDF también se puede descargar y ya estamos trabajando para que el resto también sea descargable. Obtén más información aquí.
¿En qué se diferencian los planes de precios?
Ambos planes te permiten acceder por completo a la biblioteca y a todas las funciones de Perlego. Las únicas diferencias son el precio y el período de suscripción: con el plan anual ahorrarás en torno a un 30 % en comparación con 12 meses de un plan mensual.
¿Qué es Perlego?
Somos un servicio de suscripción de libros de texto en línea que te permite acceder a toda una biblioteca en línea por menos de lo que cuesta un libro al mes. Con más de un millón de libros sobre más de 1000 categorías, ¡tenemos todo lo que necesitas! Obtén más información aquí.
¿Perlego ofrece la función de texto a voz?
Busca el símbolo de lectura en voz alta en tu próximo libro para ver si puedes escucharlo. La herramienta de lectura en voz alta lee el texto en voz alta por ti, resaltando el texto a medida que se lee. Puedes pausarla, acelerarla y ralentizarla. Obtén más información aquí.
¿Es A Practical Guide to Instrumental Analysis un PDF/ePUB en línea?
Sí, puedes acceder a A Practical Guide to Instrumental Analysis de Erno Pungor, G. Horvai en formato PDF o ePUB, así como a otros libros populares de Physical Sciences y Chemistry. Tenemos más de un millón de libros disponibles en nuestro catálogo para que explores.

Información

Editorial
CRC Press
Año
2020
ISBN
9781000142440
Edición
1
Categoría
Chemistry
Part I
Electroanalytical Techniques
Chapter
Potentiometry
1

Principle of the Technique

Techniques based on the measurement of potential of a sensor are termed potentiometry. In a Potentiometric type of sensor a membrane or sensor surface (indicator electrode) acts as a half-cell, generating a potential proportional to the logarithm of the analyte activity (concentration). The indicator electrode is connected directly (or through a salt bridge) with a reference electrode to form a galvanic cell. The electromotive force (EMF) or cell voltage of the galvanic cell thus obtained is measured in such a manner that no current flows through the cell. For a complete electrochemical cell the cell voltage is
Ecell=EindEref+Ej
where Ecell is the cell voltage or the EMF, Eind is the potential of the indicator electrode, Eref is the potential of the reference electrode, and Ej is the liquid-junction potential.
The potential of the indicator electrode is given by the following equation:
Eind = Constant +2.303RT/zF log a
where 2.303RT/zF = Nernst factor (its value at 25°C and in the case of monovalent ions is 59.16 mV), z = charge of the ion measured, and a = activity of the ion measured.
Potentiometric methods belong to two major types:
1. Direct potentiometry. Direct measurements of concentrations or activities. In this case the ion activity or ion concentration is determined by means of a calibration curve or the standard addition technique.
2. Indirect potentiometry or Potentiometric titration. The component to be determined is titrated with a suitable titrant and the indicator electrode is used to follow the changes in potential in the course of the titration.

1.1. Determination of pH in Fruit Juice

Principle of the Determination

The term pH is simply a mathematical symbol of convenience, widely accepted and firmly established for expressing the acidity of aqueous solutions. It is the negative logarithm of activity of the hydrogen ion:
pH=paH+=log aH+
where aH+ = f+[H+] (f+ = single ion activity coefficient).
However, single ion activity coefficients cannot be measured directly; only the mean activity coefficients are available (f±).
Since the pH values defined by this formula cannot be determined exactly, it was necessary to introduce the practical (“operational”) pH scale, which is based on standards. The most commonly used pH scale is the NBS (National Bureau of Standards) pH scale.
For the determination of the practical pH values, the following equation can be used:
pH(x)=pH(s)+(ExEs)/2.302RTF1
where pH(x) is the pH of the test solution and pH(s) is the pH of the standard.
Ex and Es are the EMF values measured in the test solution and in the standard, respectively. It has been suggested that in practical pH determinations with a glass electrode the best approach is to use two standards from the NBS scale, one below and one above the pH value of the test solution. The EMF of the cell containing test solution x (Ex):
glass electrode solution reference electrode
is measured and the EMF values E1 and E2 are similarly measured using the same cell containing standard solutions S1, and S2, respectively, having pH values pH1, and pH2 on either side as near as possible to the pH of x (=pHx).
(pHxpH1)/(pH2pH1)=(ExE1)/(E2E1)

Apparatus

mV- pH meter
Glass electrode
Reference electrode (Ag/AgCl or calomel)

Chemicals

pH buffers

Sample

Fruit juice

Procedure

Prepare the pH buffer solutions and measure the Emf values in the standards and the test solutions. Repeat the measurements on the pH scale after adjusting the pH meter to the pH value corresponding to the pH of one of the standards (e.g., pH = 7), then to the pH value of a second standard (e.g., pH = 2). Finally, the pH of the sample is measured.

Evaluation

Calculate the pH on the basis of the millivolt measurements and compare it with that obtained by direct pH measurement.

1.2. Determination of Fluoride in Toothpaste with Fluoride Ion-Selective Electrode Using Standard Addition

Principle of the Determination

The solid-state fluoride electrode has found extensive use in the determination of fluoride in a variety of materials. A total ionic strength adjustment buffer (TISAB) is used to adjust all unknowns and standards to essentially the same ionic strength; therefore, the concentration of fluoride can be measured and evaluated. The pH of the buffer is about 5, a level at which F is the predominant fluorine species. The buffer also containes complexing agent, which forms stable chelates with iron(III) and aluminum(III), thus eliminating their possible interference.

Solutions

1. TISAB. Buffer solution for 15 to 20 determinations can be prepared by mixing (with continuous stirring) 57 ml of glacial acetic acid, 58 g of NaCl, 4 g of cyclohexylamine dinitrilo-triacetic acid, and 500 ml distilled water in 1-1 beaker. Cool the content and add 6 M NaOH until a pH of 5.0 to 5.5 is reached. Dilute 1 1 with water and store in a plastic bottle.
2. Standard fluoride solution (2 × 10−1 mol/l). Dry a quantity of NaF at 110°C for 2 h. Cool in a desiccator, then weigh the appropriate amount of salt into a 11 volumetric flask. (Caution! NaF is highly toxic.) Dissolve in water, dilute to the mark, mix well, and store in a plastic bottle.

Apparatus

pH meter
Solid-state fluoride electrode
Reference electrode (Ag/AgCl double junction or calomel electrode)
Magnetic stirrer

Procedure

1. Determination of the slope of the fluoride calibration curve. Prepare 100, 100-ml calibration solutions in the concentration range of 10−5 to 10−2 mol/l fluoride containing fifty 50-ml TISAB buffer. From the calibration curve the (E vs. −log cF−) the slope is determined.
S=ΔE/Δ( log c)
2. Determination of fluoride in toothpaste. Weigh about 1 g of toothpaste into a 250-ml beaker. Add 50 m...

Índice

Estilos de citas para A Practical Guide to Instrumental Analysis

APA 6 Citation

Pungor, E., & Horvai, G. (2020). A Practical Guide to Instrumental Analysis (1st ed.). CRC Press. Retrieved from https://www.perlego.com/book/2039043/a-practical-guide-to-instrumental-analysis-pdf (Original work published 2020)

Chicago Citation

Pungor, Erno, and G Horvai. (2020) 2020. A Practical Guide to Instrumental Analysis. 1st ed. CRC Press. https://www.perlego.com/book/2039043/a-practical-guide-to-instrumental-analysis-pdf.

Harvard Citation

Pungor, E. and Horvai, G. (2020) A Practical Guide to Instrumental Analysis. 1st edn. CRC Press. Available at: https://www.perlego.com/book/2039043/a-practical-guide-to-instrumental-analysis-pdf (Accessed: 15 October 2022).

MLA 7 Citation

Pungor, Erno, and G Horvai. A Practical Guide to Instrumental Analysis. 1st ed. CRC Press, 2020. Web. 15 Oct. 2022.