CHAPTER ONE
The Physics of Ultrasound
Cardiac ultrasound, echocardiography, permits noninvasive and nonionizing visualization of the inside of the heart including the aorta, the ventricles and atria, the auricular appendages, and all of the cardiac valves. Dynamic images of the contracting heart are created with two-dimensional and motion mode (M-mode) images while blood flow through the heart can be seen and measured with Doppler ultrasound. Tissue Doppler imaging allows analysis of myocardial motion. Defects including valvular lesions, cardiac shunts, cardiac and thoracic masses, pleural and pericardial effusions, myocardial diseases, and stenotic lesions can be seen. More importantly it allows assessment of cardiac chamber sizes, cardiac function, blood flow, and myocardial motion, which provides information on hemodynamic status and extent of the disease process.
Uses of Echocardiography
- See internal cardiac structures.
- Evaluate function.
- Evaluate size.
- See defects.
- Valvular lesions
- Shunts
- Myocardial abnormalities
- Masses
- Effusions
- Stenotic lesions
- Evaluate blood flow.
- Assess myocardial motion and function.
All of this is possible because of sound. Sound is sent into the body and reflected from soft tissue structures. The reflected sound waves are analyzed, and an image is generated on a monitor. Sending out many sound waves side by side will produce an image with depth and width. The result is a two-dimensional image (Figure 1.1). When the sound waves are continuously and rapidly sent out in sequence, many two-dimensional images can be generated per minute, and a moving image of the heart is made called real-time or B-mode ultrasound. By sending out only one sound beam instead of many, only the structures associated with that one beam are seen, producing an M-mode image (Figure 1.2). The structures associated with that one line through the heart keep scrolling on the screen as the heart continues to contract and relax. The M-mode image displays depth on the vertical axis and time along the horizontal axis.
Doppler is used in diagnostic ultrasound to provide information on blood flow (spectral and color-flow Doppler) or myocardial motion (tissue Doppler imaging [TDI]) of the heart and its vessels. Specific locations within the heart can be selected and a spectral display of blood flow or muscle motion is created. As in M-mode the horizontal axis represents time while the vertical axis represents velocity (Figure 1.3).
This chapter deals with the physical principles of sound waves that allow ultrasound to be used as a diagnostic tool. The physics of ultrasound involves an understanding of the basic properties of sound waves and how these properties affect transducer selection, image quality, and diagnostic interpretation. Only the principles needed to make knowledgeable technical decisions and diagnostic interpretations are presented in this chapter. More detailed information can be found in books dedicated to the physics of diagnostic ultrasound. Selected references are listed at the end of the chapter.
Basic Physics
Cycles and Wavelengths
Sound waves travel in longitudinal lines within a medium. The molecules along that longitudinal course of movement are alternately compressed (molecules move closer together) and rarefacted (molecules are spread apart). The time required for one complete compression and rarefaction to occur is one cycle (Figure 1.4). The distance in millimeters that the sound wave travels during one cycle is its wavelength.
Sound Waves
- Alternately compress and spread apart the molecules in their pathway.
- 1 cycle = one complete compression and expansion
- Wavelength = the distance traveled during 1 cycle
The source of the sound wave determines the length of a cycle. Transducers generate the sound in diagnostic ultrasound. They will be discussed in detail later, but for any given transducer the wavelength is constant.
Frequency
The number of cycles per second is the frequency of the sound wave (Figure 1.5). Frequency is measured in Hertz (Hz), where 1 Hz equals one cycle per second. Ultrasound has a frequency greater than 20,000 cycles per second, and is beyond the range of human hearing. Since frequency is the number of complete cycles per second, the higher the frequency of the sound wave the shorter the wavelength must be.
Frequency
- The number of cycles per second = frequency
- High frequency = shorter wavelengths
- Low frequency = longer wavelengths
A 5.0-megahertz (MHz) transducer transmits 5 million cycles per second at 0.31 millimeters (mm) per cycle, while a 2.0-MHz transducer transmits only 2 million cycles per second at 0.77 mm per cycle. Table 1.1 lists wavelengths for sound generated at various frequencies.
Table 1.1 Wavelength of Sound at Commonly Used Frequencies
Speed of Sound
The speed of sound (V) depends upon the density and stiffness of the medium through which it is traveling. Increased density allows sound to travel faster. The velocity of sound does not change within a homogeneous substance and is independent of frequency (Figure 1.6). Table 1.2 lists the speed of sound in various tissues. The speed of sound through air is very slow because of its low density, while bone allows sound to travel at relatively high speeds.
Table 1.2 The Speed of Sound in Soft Tissues
|
air | 330 |
fat | 1,440 |
brain | 1,510 |
liver | 1,560 |
kidney | 1,560 |
muscle | 1,570 |
blood | 1,570 |
bone | 4,080 |
The average velocity of a sound wave in soft tissue is 1,540 meters per second regardless of transducer frequency (Figure 1.7). Velocity is calibrated into the ultrasound machine, which then calculates the distance (D) to cardiac structures based upon how long it takes to receive reflected echoes (T):
Transducer frequency does not affect the speed of sound in tissues.
The time (T) required to travel 1 cm is 6.5 micr...