MRI PHYSICS

MRI PHYSICS

TECH TO TECH EXPLANATIONS

Technologists must have a solid understanding of the physics behind Magnetic Resonance Imaging (MRI), including safety, the hows and whys of the quantum physics of the MR phenomenon, and how to competently operate MRI scanners. Generating the highest quality images of the human body involves thorough knowledge of scanner hardware, pulse sequences, image contrast, geometric parameters, and tissue suppression techniques.

MRI Physics: Tech to Tech Explanations is designed to help student MRI technologists and radiotherapists preparing for Advanced MRI certification examinations to better understand difficult concepts and topics in a quick and easy manner.

Written by a highly experienced technologist, this useful guide provides clear and reader-friendly coverage of what every MR Technologist needs to know. Topics include safety considerations associated with the magnetic field and RF, pulse sequences, artifacts, MRI math, the much-feared gradients, and I.V. contrast.

  • Provides basic guidance on safety considerations, protocols options, critical thinking, and image contrast optimization
  • Simplifies the challenging topic of MRI physics using straightforward language and clear explanations
  • Covers content for American Registry of Radiologic Technologists (ARRT) and Continuing Qualifications Requirements (CQR) exams
  • Features numerous illustrations and photographs of various MRI concepts, pulse sequence design, artifacts, and the application of concepts in clinical settings

MRI Physics: Tech to Tech Explanations is a must-have resource for the experienced and training MRI technologist, medical students, and radiology residency rotations.

Häufig gestellte Fragen

Wie kann ich mein Abo kündigen?
Gehe einfach zum Kontobereich in den Einstellungen und klicke auf „Abo kündigen“ – ganz einfach. Nachdem du gekündigt hast, bleibt deine Mitgliedschaft für den verbleibenden Abozeitraum, den du bereits bezahlt hast, aktiv. Mehr Informationen hier.
(Wie) Kann ich Bücher herunterladen?
Derzeit stehen all unsere auf Mobilgeräte reagierenden ePub-Bücher zum Download über die App zur Verfügung. Die meisten unserer PDFs stehen ebenfalls zum Download bereit; wir arbeiten daran, auch die übrigen PDFs zum Download anzubieten, bei denen dies aktuell noch nicht möglich ist. Weitere Informationen hier.
Welcher Unterschied besteht bei den Preisen zwischen den Aboplänen?
Mit beiden Aboplänen erhältst du vollen Zugang zur Bibliothek und allen Funktionen von Perlego. Die einzigen Unterschiede bestehen im Preis und dem Abozeitraum: Mit dem Jahresabo sparst du auf 12 Monate gerechnet im Vergleich zum Monatsabo rund 30 %.
Was ist Perlego?
Wir sind ein Online-Abodienst für Lehrbücher, bei dem du für weniger als den Preis eines einzelnen Buches pro Monat Zugang zu einer ganzen Online-Bibliothek erhältst. Mit über 1 Million Büchern zu über 1.000 verschiedenen Themen haben wir bestimmt alles, was du brauchst! Weitere Informationen hier.
Unterstützt Perlego Text-zu-Sprache?
Achte auf das Symbol zum Vorlesen in deinem nächsten Buch, um zu sehen, ob du es dir auch anhören kannst. Bei diesem Tool wird dir Text laut vorgelesen, wobei der Text beim Vorlesen auch grafisch hervorgehoben wird. Du kannst das Vorlesen jederzeit anhalten, beschleunigen und verlangsamen. Weitere Informationen hier.
Ist MRI Physics als Online-PDF/ePub verfügbar?
Ja, du hast Zugang zu MRI Physics von Stephen J. Powers im PDF- und/oder ePub-Format sowie zu anderen beliebten Büchern aus Médecine & Technologie et fournitures médicales. Aus unserem Katalog stehen dir über 1 Million Bücher zur Verfügung.

Information

Jahr
2021
ISBN
9781119615040

1
Hardware: Magnet Types and Coils

Magnets

Three types of magnets are used in MRI:
  • Permanent. These are rare to almost nonexistent today. They are made up of brick‐sized magnets that, when placed in the right pattern, combine to make a magnetic field strong enough to image with. These magnets are large, heavy, and have a weak B0 field so signal‐to‐noise ratio (SNR) is at a premium. They cannot however be turned off.
  • Resistive. These consist of a large coil of wire with a lot of electricity passing through it. The amount of electricity circulating is sufficient to make a magnetic field strong enough to image with. These are usually 0.5–0.7 T. They also generate a lot of heat from electrical resistance in the wire.
  • Superconducting. These are resistive magnets but with one major difference. The magnet is cooled with liquid helium, a cryogen. The cryogen decreases electrical resistance in the wires that make up the magnet. The loss of resistance from cooling enables more current to flow so stronger magnetic fields are possible. Typical field strengths are 1–4T. Research magnets can be as strong as 7–9 T. Stronger magnetic fields also come with a larger fringe field compared to the lower‐power fields. Think of the fringe field as being like the “Scatter Radiation” in x‐radiography. The magnetic field is not solely confined to the scanner. Magnetism picks up rapidly as you get closer to the bore.
The field attraction gets stronger the closer you get to isocenter.

Teaching Moment:

Magnetic force of attraction is a cubic exponential function. Halve your distance and the strength of the fringe field you are experiencing goes up by a factor of 8 (23, or 2 × 2 × 2). That is why pens, beepers, and paper clips get pulled out of your pockets and go flying into the bore at about 40 mph.

The Superconducting Magnet

The magnet coil is cooled with liquid helium. Helium is usually a gas, but when it is sufficiently compressed it becomes a liquid which is extremely cold. Liquid helium is a cryogen with a temperature of about 4° Kelvin (very close to absolute zero).
Helium does not want to be a liquid. It wants to be a gas and revert to a lower energy state. If allowed to, it will do so very rapidly. The resulting conversion back to a gas is called a quench. Quenches result in a high conversion ratio of about 750:1, meaning that 1 litre of liquid helium makes about 750 litres of gaseous helium. Superconducting magnets have a large vent pipe leading outside the building so that if there is a quench, the gases vent outside and do not fill the scan room.
Quenching the magnet is only done in a life‐threatening situation such as when someone is pinned between a ferrous object and the magnet. Trust me, if an O2 bottle, IV pole, or stretcher is drawn in to the magnet, you will NOT get it off. A quench will result in the scanner being down for several days.

Coils

There is a plethora of different coils. Some are general purpose; others are highly specific. General purpose or flexible coils are made by multiple manufacturers. They can be used for a shoulder on one patient, a knee on the next, then a soft tissue mass on another.
Coils for more specific use are breast, endo‐rectal, or the long‐lost Temporal‐Mandibular Joint (TMJ) coils. If you still have TMJ coils and they still work on your system, keep them, they are awesome for wrists, orbits, and fingers! Remember, a coil is a coil is a coil. But also know that coils are system specific. You cannot use a 1.5 T coil on a 3 T system and vice versa.
Coils vary in size: small, medium, and large. If you can get it around, under, or over the area to be imaged, it will work. However, you have to choose a coil size that is appropriate to the patient size or coverage needed. You should use the smallest coil possible for best signal to noise. With all scan factors remaining the same, a small coil has better SNR than a large coil. Why? The smaller coil unto itself does not see more signal, it sees less noise. That’s the signal to noise ratio. The same amount of signal with less noise results in a higher SNR.
An imaging coil is just that: loops of wire that when placed next to or around a body part form an integral part of something in MR called Faraday’s law of induction. Faraday’s law of induction states that a moving magnet field passing through a coil of wire will induce current to flow in that wire. Current is signal, and signal is money.
Schematic illustration of the four basic coil types.
Figure 1.1 The four basic coil types. Typical body parts they are used on are:
  • Surface coil: spine
  • Saddle coil: knee, foot/ankle
  • Helmholtz: shoulder (Figure 1.2)
  • Bird cage (quadrature): brain (Figure 1.3).
Figure 1.1 shows the four basic coil types. Remember, t...